Model EMPX-HF (Lakeshore Cryogenics)
The cryogenic probe station is used to make electical measurements on devices under a controlled temperature and magnetic field. Up to four DC or RF probes can be connected simultaneously.
Manufacturer's Website
Capabilities
-
360 degree in-field sample stage rotation
-
1" grounded sample chuck
-
Probe travel distance of 2" on X axis, 1" on Y axis, and 0.71" on Z axis
-
Measurements up to 40 GHz with GSG microwave probes
-
Operating temperature range from 5 K to 400 K
-
Temperature control stability with LHe:
-
Better than 5 mK RMS from 5 K to 50 K
-
Better than 50 mK RMS above 50 K
-
Temperature control stability with LN2 better than 100 mK RMS
-
Horizontal magnetic fields up to 550 mT
-
Magnetic field uniformity:
-
0.6% over 1 cm
-
1.1% over 1.27 cm
-
2% over 2 cm
-
2.6% over 2.54 cm
-
Field control stability better than 5 µT RMS